Aid 4D probe; realize the probe by means of the ID code; swap the 4D generate signal by using the  electrical relay to generate the 4D probe operate Usually. 4.2.2 Key board The key board can be divided into two elements: front-conclude of the most crucial board and again-stop of the primary board.

Cookie penting untuk memfungsikan situs dengan semestinya. Untuk meningkatkan pengalaman Anda, kami menggunakan cookie untuk mengingat element login Anda dan menyediakan login aman, mengumpulkan statistik untuk mengoptimalkan fungsionalitas situs, dan menyediakan konten yang disesuaikan dengan minat Anda.

Observe: On the check, the brightness adjustment arrives right before distinction. Following readjusting the watch’s distinction and brightness, change all preset and peripheral settings.

Don't use any cables in addition to the cables delivered Using the product by Mindray. 3. Use the cable delivered with this system to attach the printer. Other cables may perhaps lead to electrical shock. Website page fifteen: Mechanical Security

Slash off cable ties connecting the principle unit about the Screen by diagonal reducing pliers, after which pull out the corresponding cable plugs.

Information Strategy Examine the battery periodically : a) Verify if battery can be billed Commonly when ability-on: That the current ability is 100% or capability increases just after a short time implies that the battery is usually charged Generally.

The chapter materials the tactic to confirm principal perform and effectiveness of products. This is often only employed for reference, not preventive execution. Operate examining and screening of the part shall be performed by Mindray provider engineers and the consumer collectively. Web page sixty six: Basic Test

Klik Setuju dan Lanjutkan untuk menyetujui cookie dan langsung menuju ke situs atau klik Informasi Selengkapnya untuk melihat deskripsi depth mengenai tipe cookie dan memilih apakah akan menyetujui cookie tertentu saat berada di situs ini.

Permits utilization of many scanned angles to variety just one picture, leading to Improved contrast resolution and improved visualization

five.2.three Geometric positioning accuracy Longitudinal geometric positioning precision  Take a look at Phase: Do changes as the way in tests the utmost depth. Record the gap by 20mm Every single segment to the longitudinal targets line using the measurement caliper; Pick out the value with the best error (to 20mm), work out the accuracy utilizing the formulation beneath The measurement caliper ought to be positioned at the upper fringe of the focus on, Take note:... Page seventy nine Transverse geometric positioning precision  Check Action: Deal with the scan surface area of the phantom with drinking water or couple gel, gently Get hold of the probe Together with the scan surface area Alter the depth, generating the transverse targets to become displayed during the impression. Alter the focus level to become posited beside the transverse targets (the conventional is just not very clear) Alter parameters like gain, TGC, making Every transverse targets for being clearly displayed. Web page 80 five.five.2.4 Blackout Location Check Phase: Cover the scan area of your phantom with h2o or pair gel, Carefully Make contact with the probe with the scan Dp50rb area Change the depth at a reduce value, and set the main target at the closest spot to the scan surface area. Reduce the worth of parameters like AP, Attain until eventually the track record sounds just is often witnessed. Website page eighty one: Software Upgrade And Upkeep

Kebutuhan Anda akan alat profesional dan canggih yang dikombinasikan dengan pencitraan berkualitas dan kemudahan kegunaan kini dapat dipenuhi dengan sistem portabel baru Mindray DP-50Exp.

Explore greater diagnostic information via extended watch on the anatomical structure on all linear probes

Overall performance Exam 5.5.1 Examination Process Resolution Detection depth Geometric Placement Precision Blackout space check Report the Test photos 5.5.2 Check Information The picture utilized Here's just for reference, keep on with the image outcome in the actual Take note: situation. Specifications: Screen: set the distinction and brightness at the scientific software worth (or even the default status) Procedure ecosystem: dim area, simulating the clinical software natural environment. Web site 76 In problem that the transverse resolution screening targets are horizontally displayed, record the minimal length of two targets that may be clearly identified. Repeat the operation higher than for that transverse resolution testing targets at other depths. As demonstrated in figure underneath. Axial resolution ... Webpage seventy seven When using the convex probe, continue to keep the transverse resolution screening targets to notice: be displayed near the midline. When utilizing a linear probe with steer function, will not turn on the steer purpose when carry out the transverse resolution take a look at. Zoom inside the area where the targets Positioned if necessary. The diameter on the concentrate on stage at a particular depth is equal into the transverse resolution in the depth. Website page seventy eight five.

Following replacing the CPU module, key file have to be created once again based on the new MAC tackle and written into HDD on the equipment. Or else, the configuration function in the product or service will likely be missing. After eliminating the CPU module, Clear away panhead screws with washers (twelve M3X6) fastened on the main device box, you may take from the board. Page 118 Determine eight-29 Disassemble the Control Panel Assembly (1) Keep the higher edge of control panel assembly, and open up up the control panel. Figure eight-30 Disassemble the Control Panel Assembly (2) Following getting rid of a single countersunk head self-tapping screw PT3X10 set around the user interface and get out the plug in the keyboard and the key device connecting cables, you could take out the user interface assembly.

Notice: Ensure that you could potentially pull out and drive in the IO board sockets vertically when pulling out and assembling the IO board assembly on the placement with the handle. Get rid of panhead screws with washers M3X6 (four) utilized to protected the IO board, after which you could possibly choose out the IO board. Web page 116 Get rid of panhead screws with washers (2 M3X6) that happen to be fastened on the leading device, acquire out the upper baffle then pull out two cables connecting the leading board to the ability module assembly.